Specifications for DRE Waveline Pro Multi-function Veterinary Monitor
Physical Dimensions & Weight:
- Base Unit: 12.1”/309 mm (H) x 14.4”/365 mm (W) x 6.3”/159 mm (D)
- Weight: 17.6 lbs/8.0 kgs (including anesthetic agents module and batteries)
Application:
- Small, Medium, and Large animal
Performance Specifications:
- Display: 15” color TFT, Resolution 1024 x 768
- Trace: 8 waveforms
- Indicators:
- Alarm indicator
- Power indicator
- QRS beep and alarm sound
- Trend time: 1 – 48 hours
- Recorder:
- Built-in, thermal array, 2 channels
- Record width: 1.9”/48 mm
- Recorder paper: 2”/50 mm
- Record speed: 25 mm/s, 50 mm/s
ECG:
- Input: 5-lead ECG cable and standard AAMI line for connection
- Lead Choice: I, II, III, aVR, aVF, aVL, V, TEST
- Gain Choice : x0.5, x1, x2, x4
- Frequency Characteristic: 0.05 ~ 35 HZ (+3dB)
- ECG Waveforms: 7 channels
- Penetration Voltage: 4000VAC 50/60Hz
- Sweep Speed: 12.5, 25 and 50 mm/s
- HR Display Range: 30 – 300bpm
- Accuracy: 1 bpm or 1%, whichever is greater
- Alarm Limit Range Setting: Upper limit 100 – 200 bpm; Lower limit 30 – 100 bpm
RESP:
- Measure Method: RA-LL impedance
- Range: 0 – 120 rpm
- Accuracy: 3 rpm
- Alarm Limit Setting: Upper limit 6 – 120 rpm: Lower limit 3 – 120 rpm
- Sweep Speed: 12.5 and 25 mm/s
TEMP:
- Range: 25 – 50 (C)
- Accuracy:
- 0.2 C (25.0 – 34.9 C)
- 0.1 C (35.0 – 39.9 C)
- 0.2 C (40.0 – 44.9 C)
- 0.3 C (45.0 – 50.0 C)
- Display Resolution: 0.1 C
- Alarm Limit Setting: Upper limit 0 – 50 C, lower limit 0 – 50 C
- Channel: 2 channels
Masimo SET SpO2
- Measuring Range:
- SpO2: 1~100 %
- PR: 25~20 %
- SpO2 Accuracy (non-motion):
- Adult Pediatric: 70~69%: unspecified
- Neonate: 70~69 %: unspecified
- SpO2 Accuracy (motion):
- Adult Pediatric: 70~100 %: ±3 %, 0~100 %: ± 3 %, 0~240 bpm: ±3 bpm
- PR Accuracy (motion):
- Adult Pediatric Neonate: 25<span 20.8px;"="">~240 bpm: ±5 bpm
- Low Perfusion Performance: >0.02 % Pulse Amplitude and % Transmission >5
- SpO2: ±2 %
- PR: ±3 bpm
- Modes:
- Averaging mode: 2,4,8,10,12,14 and 16 s
- Sensitivity: Normal, APOD and Maximum
NIBP:
- Measuring Technology: Automatic oscillating measurement
- Cuff Inflating: <30s (0 – 300 mmHg, standard adult cuff)
- Measuring Period: AVE<40s
- Mode: Manual, Auto
- Measuring Interval in AUTO Mode: 2 min – 4 hrs
- Pulse Rate Range: 30 – 250 (bpm)
- Measuring Range:
- Adult/Pediatric Mode: SYS: 40 – 250 (mmHg) DIA :15 – 200 (mmHg)
- Neonatal Mode: SYS: 40 – 135 (mmHg) DIA : 15 – 100 (mmHg)
- Resolution: 1mmHg
Accuracy:
- Maximum Mean error: 5mmHg
- Maximum Standard deviation: 8mmHg
- Overpressure Protection:
- Adult Mode: 300(mmHg)
- Neonatal Mode: 160 (mmHg)
- Alarm Limit Setting:
- SYS: 50 – 240 mmHg
- DIA: 15 – 180 mmHg
IBP:
- Measurement Range: -50 – 300mmHg
- Channel: 2 channels
- Pressure Transducer: Sensitivity, 5μV/V/mmHg
- Impedance Range: 300 – 3000Ω
- Transducer Sites: ART, PA,CVP, RAP, LAP, ICP
- Unit: mmHg/kPa selectable
- Resolution: 1 mmHg
- Accuracy: 1 mmHg or 2%, whichever is greater
- Alarm Range: -10 – 300 mmHg
EtCO2 (Respironics) (mainstream or sidestream):
- CO2 Measurement Range: 0 – 99 mmHg
Accuracy:
- 2 mmHg (0 – 38 mmHg)
- 39-99 mmHg 5% of reading + 0.08% for every 1 mmHg (above 38 mmHg)
- Sampling Rate: 50 ml/min
- Initialization Time: 30 seconds (typical), reaches 5% steady-state accuracy within 3 minutes
- Respiration Rate: 0 – 150 breaths/min
- Mode: Adult, neonate
Anesthetic Agents:
- Industry standard 802.11b/g wireless network
Power:
- Source: External AC power or internal battery
- AC Power: 100 – 240VAC, 50/60Hz, 150VA
- Battery: Built-in and rechargeable
- Charge Time: 4 hours
Environmental Specifications:
- Operating Temperature: 5 – 40 C
- Storage Temperature: -20 – 65 C
- Operating Humidity: ≤80 %
- Storage Humditiy: ≤80 %