Specifications for DRE Waveline Touch Veterinary Monitor
Physical Dimensions
- 21.9” H x 13.2” W x 9.23” D (555 mm H x 335 mm W x 235 mm D)
- Weight - 11bs (5kg)
Application
- Veterinary patients
Power
- Power Source - External AC power or internal battery
- AC Power - 100-240VAC, 50/60Hz, <150VA
- Battery - Rechargeable Lead-Acid; Operating time under the normal condition (1 hour)
- Operating time after the first alarm of low battery (10 min.)
Environmental
- Working Temperature: 5-40 C
- Storage Temperature: 0-65 C
- Working Humidity: 30-75%
- Storage Temeprature: ≤80%
- Operating Altitude: Up to 5,000 meters
- Hyperbaric Pressure - Up to 405.3 kPa
Performance Specifications
- Display - 12.1” color TFT, Touch Screen
- Resolution - 800 x 600 pixels
- Trace - 4, 6 or 8 waveforms ECG (I, II, III, aVR, aVL, aVF, V1-V6), PLETH, RESP, IBPx2, ETCO2
- Indicator - Alarm indicator; Power indicator QRS beep and alarm sound
- Trend Time - 72 hours
- Recorder - Built-in, thermal array, 2 channels
- Record - Width 48mm
- Recorder - Paper 50mm
- Record - Speed 25mm/s, 50mm/s
- OxyCRG - Combines HR, SpO2, and resp. trends into a single graph ECG
- Input - 5-lead ECG cable and standard AAMI line for connection
- Lead Choice - I, II, III, aVR, aVF, aVL, V
- Gain Choice - x0.125, x0.25, x0.5, x1.0, x2.0
Filter
- Diagnostic mode - 0.05 – 100Hz
- Monitoring mode - 0.05 – 75Hz
- Surgical mode - 1 – 20Hz
- Frequency Characteristic - 0.05-40 Hz
- ECG Waveforms - 7 channels
- Penetration Voltage - 4000VAC 50/60Hz
- Sweep Speed - 12.5, 25 and 50 mm/s
- HR Display Range - 30-300bpm
- Accuracy - 1% or 1bpm, whichever is greater
- S-7 Segment Detection - 7 channels
- Measurement Range - -2.0mv – +2.0mv
- Arrhythmia Analysis - 13 types
- Alarm Limit Range Setting - Upper limit 80-400bpm; Lower limit 20-150bpm
RESP
- Measure Method - RA-LL impedance
- Range - 0-120 rpm
- Accuracy - + 3 rpm
- Alarm Upper-lower Limit Setting - Upper limit 6-120 rpm, Lower limit 3-120 rpm
- Sweep Speed - 12.5 and 25 mm/s
NIBP
- Measuring Technology - Automatic oscillating measurement
- Cuff Inflating - <30s (0 – 300mmHg, standard adult cuff
- Measuring Period - AVE<40s
- Mode - Manual, Auto, STAT
- Measuring Interval in AUTO Mode - 2 min-4 hrs
- Pulse Rate Range - 30-250 (bpm)
- Measuring Range - Adult/Pediatric Mode, SYS 40-250 (mmHg), DIA 15-200 (mmHg); Neonatal Mode, SYS 40-135 (mmHg), DIA 15-100 (mmHg)
- Resolution - 1mmHg
- Accuracy - Pressure, Maximum Mean error 5mmHg, Max. Standard deviation 8mmHg
- Overpressure Protection - Adult Mode 280(mmHg), Neonatal Mode 150 (mmHg)
- Alarm Limit Setting - SYS 50-240 mmHg, DIA 15-180 mmHg
TEMP
- Range - 0 – 50 C (32 – 122 F)
- Accuracy - Without sensor
- Display Resolution - 0.1 C
- Alarm Upper-lower Limit Setting - Upper limit 0-50 C, Lower limit 0-50 C
- Channel - 2 channels, provide T1, T2, ΔT
- Alarm Limit Setting - 10-50 C
Masimo SET SpO2
- Measuring Range:
- SpO2: 1~100 %
- PR: 25~20 %
- SpO2 Accuracy (non-motion):
- Adult Pediatric: 70~69%: unspecified
- Neonate: 70~69 %: unspecified
- SpO2 Accuracy (motion):
- Adult Pediatric: 70~69% : unspecified
- Neonate: 70~69 % : unspecified
- PR Accuracy (non-motion):
- Adult Pediatric Neonate: 25?240 bpm: ±3 bpm
- PR Accuracy (motion):
- Adult Pediatric Neonate: 25<span 20.8px;"="">~240 bpm: ±5 bpm
- Low Perfusion Performance: >0.02 % Pulse Amplitude and % Transmission >5
- SpO2: ±2 %
- PR: ±3 bpm
- Modes:
- Averaging mode: 2,4,8,10,12,14 and 16 s
- Sensitivity: Normal, APOD and Maximum
IBP (Option)
- Measurement Range - -10-300mmHg
- Channel - 2 channels
- Pressure Transducer Sensitivity, 5μV/V/mmHg
- Impedance Range - 300-3000Ω
- Transducer Sites - ART, PA, CVP, RAP, LAP, ICP
- Resolution - 1mmHg
- Accuracy - 1mmHg or 2%, whichever is greater
- Alarm range - -10-300mmHg
EtCO2 (Option) (Respironics)
- Mode of Sampling - Sidestream
- Principle of Operation -Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts
- CO2 Measurement Range - 0 to 150 mmHg (0 to 19.7%, 0 to 20 kPa)
- CO2 Calculation Method - BTPS (Body Temperature Pressure Saturated)
- CO2 Resolution - 0.1mmHg(0-69mmHg); 0.25mmHg(70-150mmHg)
- CO2 Accuracy - 0-40 mmHg 2 mmHg; 41-70 mmHg 5% of reading; 71-100 mmHg 8% of reading; 101-150 mmHg 10% of reading; Above 80 breath per minute 12% of reading
- Sampling Rate - 100Hz
- Respiration Rate - 2-150 bpm
- Respiration Rate Accuracy - 1 breath
- Response Time - <3 seconds - includes transport time and rise time
- Inspired CO2 Measurement Range - 3-50 mmHg
Documents: