Specifications for DRE Waveline EZ Portable Veterinary Monitor
Physical Dimensions:
- Size:6 lbs
- Dimensions:10x8x4 (wxhxd)
Safety Approval & Quality System:
- Designed to meet IEC60601-1-1988, EN60601-1-1, EN60601-2
- Class II Equipment, double insulated
- Type BF applied parts
- ISO9001 & EN46001 Certified
Power Requirements:
- Power Supply: AC 90-264V/47-63Hz
- Input Power: ≤55VA
- Fuses: Two fuse sockets in the rear panel indicated by “FUSE”, OI 5X20, 2A/250V
- Battery: 12V/4.0AH sealed lead-acid
- Charge time: ≥4 hours
- Operating time: ≥2 hours (full recharge)
- Battery Charging Method: Automatic charging after monitor is connected to AC power supply (with charge protection function)
- Discharge Protection: When powered by battery, the monitor will be automatically turned off when battery power is almost used up
ECG:
- Patient Safety Standard: IEC60601-1-1988
- CMRR ≥60dB (Common Mode Rejection Ratio)
- Heart Rate Range: 20 ~ 254bpm 1bpm
- Heart Rate Averaging: 8 second average
- ST Segment Range: -0.8 ~ + 0.8mV
- Interface: AAMI 6-pin
- Lead Selection: I, II, III (3 lead mode)
- I, II, III, aVR, aVL, aVF, V (5 lead mode) (ST and Arrhythmia analysis)
- Lead Fault Alarm: Audible, Visual Input 5-lead ECG patient cable
- QRS Indicator: Audible and Visual Alert
- Waveform Storage: 6 minutes
- Sweep Speed: 12.5/25/50 mm/sec
- Gain Selection:4mV, 2mV, 1mV, 0.5mV, 0.25mV,Auto
- Trends: 2 hours -> 4 hours -> 8 hours -> 24 hours -> 48 hours
- Patient Isolation Breakdown: voltage 4000VAC 50Hz 60 seconds Leakage current <10μA
- Frequency width Monitoring: mode 0.5 ~ 40Hz (+0.4dB,-3.0dB) Surgery mode 0.5 ~ 20Hz (+0.4dB, -3.0dB) not calibration significant
- Patient Drive Current: <10μA
- Enclosure Leakage Current: <0.1mA
- Maximum T Wave Rejection Capability: 1.2mV
- Heart Rate Alarm Response Time:< 7 seconds
- Aspect Ratio: 0.24 ~ 0.6 sec/mV
- Alarm Frequency Low alarm: 2-2.4kHz High alarm: 3-3.4kHz
- Defibrillator Protected & ESIS Protected Tested with 5kV
- Recovery Time Following Defibrillation: <5 seconds
RESP:
- Measurement Method: Thoracic Impedance
- Respiration Rate Range: 0 ~ 1001rpm
- Accuracy: 2 rpm
Masimo SET SpO2
- Measuring Range:
- SpO2: 1?100 %
- PR: 25?240 bpm
- Perfusion: 0.02?20 %
- SpO2 Accuracy (non-motion):
- Adult Pediatric: 70?100 %: ±2 %, 0?69%: unspecified
- Neonate: 70?100 %: ± 3 %,0?69 %: unspecified
- SpO2 Accuracy (motion):
- Adult Pediatric: 70?100 %: ±3 %, 0?69% : unspecified
- Neonate: 70?100 %: ± 3 %, 0?69 % : unspecified
- PR Accuracy (non-motion):
- Adult Pediatric Neonate: 25?240 bpm: ±3 bpm
- PR Accuracy (motion):
- Adult Pediatric Neonate: 25?240 bpm: ±5 bpm
- Low Perfusion Performance: >0.02 % Pulse Amplitude and % Transmission >5
- SpO2: ±2 %
- PR: ±3 bpm
- Modes:
- Averaging mode: 2,4,8,10,12,14 and 16 s
- Sensitivity: Normal, APOD and Maximum
NIBP:
- Method Automatic oscillometric
- Parameters Systolic, diastolic, mean arterial pressure, pulse
- Scale mmHg or kPa
- Operating Modes Manual, Automatic, Continuous
- Repeat Cycles 1 ~ 10, 15, 30, 60, 90, 120 minutes
- Determination Systolic, Adult/pediatric 40 ~ 250mmHg (5.3 ~ 33.3kPa) Systolic, Neonate 20 ~ 160mmHg (2.7 ~ 21.3kPa) Diastolic, Adult/pediatric 10 ~ 180mmHg (1.3 ~ 24.0kPa) Diastolic, Neonate 10 ~ 140mmHg (1.3 ~ 18.7kPa)
- Cuff Pressure Range Adult/pediatric 0 ~ 300mmHg (0 ~ 40.0kPa) Neonate 0 ~ 140mmHg (0 ~ 18.7kPa)
- Initial Cuff Inflation Adult/pediatric 17010mmHg (22.7 1.3kPa) Neonate 10010mmHg (16.0 1.3kPa)
- Deflation Pressure 30mmHg(4.0kPa) higher than the last systolic pressure
- Cuff Inflation Rate No greater than 50mmHg/sec
- Measurement Time Typical 25 seconds, Maximum 40 seconds, Typical Stat 20 seconds
- Pressure Display Accuracy 3mmHg
- BP Pulse Rate Accuracy 2% @ 40 ~240bpm
- Cuff Neonate, infant, pediatric,standard adult
TEMP (Dual Channel)
- Range 0 ~ 50C
- Probe YSI 400 Skin surface or rectal /esophageal
- Scale Celsius
- Accuracy 0.1 C
- Resolution 0.1 C
CO2 (Respironics) (mainstream or sidestream)
- Type Side stream, mom-dispersive IR
- CO2 Range 0-99mmHg
- Scale mmHg/kPa
- Accuracy +- 2mmHg (0-40mmHg) +- 5mmHg (41-76mmHg) +-10mmHg (77-99mmHg)
- Calibration Automatic
- Respiration Range 0-150rpm, +- 2rpm
TFT Color Display
- Size 8 inches
- Matrix 640 (H) x 480 (V) pixels
- Recorder (Optional)
- Type Built-in 2-channel thermal array recorder
- Print mode Text or waveform
- Waveforms Real time or alarm-triggered
- Resolution 400dpi vertical, 800dpi horizontal
- Annotations Time, date, vital sign readings